Pulsed Discharge Detector
Model D-4-I-SH14-R
Instruction Manual
Zobrazit stránku 0
1 2 3 4 5 6 ... 22 23

Shrnutí obsahu

Strany 1 - Instruction Manual

Valco Instruments Co. Inc.Valco Instruments Co. Inc.800 · 367· 8424 sales713 · 688· 9345 tech713 · 688· 8106 [email protected] AG InternationalSch

Strany 2

6InstallationGeneral Precautions • Do not use plastic/polymer or copper tubes for gas handling andinterconnectons. Use only stainless steel tubing w

Strany 3 - Table of Contents

7InstallationGAS CHROMATOGRAPHTEE(ZT1)HELIUMPURIFIERDISCHARGE GAS(99.999% He)TGA-R-30F60PRESTRICTOR(30 mL/min minimum)VENTCOLUMNINLETDISCHARGE GASINLE

Strany 4

8InstallationEZR217. Open the cylinder valve to pressurize the regulator once again. Close thevalve and observe the needle of the high pressure gauge

Strany 5 - Introduction

9Pulser Module Installation1. Mount the pulser module as shown above in Figure 5.2. Power to the pulser module is provided through the pulser power su

Strany 6 - Safety Notes and Information

10Detector Electrical ConnectionsCAUTION: Do not use a wrench to tighten the SMCconnectors on the bias and electrometer cables.Connections should be f

Strany 7

113. Set the detector temperature to 100°C and allow time for the detector andhelium purifier to reach the set temperature.4. Set the range to 102 an

Strany 8 - System Requirements

12InstallationCapillary Column ConnectionIf the capillary column adapter is installed in the column inlet:1. Make a mark on the column 11.4 cm from th

Strany 9 - Gas Specifications

133. Set aside the 1/8" wrench and completely remove the adapter from thecolumn inlet.4. Screw the packed column adapter into the column inlet by

Strany 10 - Installation

14Mode Selection and SetupHelium Ionization ModeIf the instructions of Step 1 at the top of page 9 were properly executed,the column should already be

Strany 11 - Gas Connections

15Mode Selection and SetupAr-PDPIDChanging the discharge gas from helium to a mixture of 2% argon inhelium changes the photon energy level from the 17

Strany 12

This page intentionally left blank for printing purposes

Strany 13 - Pulser Module Installation

16TroubleshootingHigh Background CurrentLow SensitivityPurple/bluePinkDisconnect the columnand cap the column inletBaseline current is still high(>

Strany 14 - Initial Power-Up

17High Noise LevelIf the noise lievel is high:1. See if it improves with the GC fan turned off.2. Check the column insertion position (11.4 cm for cap

Strany 15

18WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Strany 16 - Packed Column Connection

19Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up,page 10), the standing

Strany 17 - Testing for Leaks

Table of ContentsIntroductionDescription and Operating Principles ... 1Safety Notes and Infor

Strany 18 - Mode Selection and Setup

This page intentionally left blank for printing purposes

Strany 19

1IntroductionDescription and Operating PrincipleThe PD-D-4-I-SH14-R is a non-radioactive pulsed discharge ionization detector(PDID) which is optimized

Strany 20 - Troubleshooting

2Helium Ionization (PDHID)The PDHID is essentially non-destructive (0.01 - 0.1% ionization) and highlysensitive. The response to organic compounds is

Strany 21 - High Noise Level

3Components of the Detector SystemComponents of the detector system are listed in the Table 1. Check thecontents of the packages to verify that every

Strany 22 - Warranty

4IntroductionSystem RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high puri

Strany 23 - Detector Performance Log

5IntroductionedoMrotceteDDIHDPDIPDP-rADIPDP-rKDIPDP-eXsagegrahcsiDmuileHeHnirA%2eHnirK%5.1eHnieX%8.0sagreirraCmuileH*****Gas Specifications** Any gas